HospitalInspections.org

Bringing transparency to federal inspections

303 N JACKSON STREET

MORRISON, IL 61270

No Description Available

Tag No.: K0012

Based on random observation during the survey walk-through, staff interview and review of Life Safety Plans dated 10-15-10 produced by Shive-Hattery Architects in Moline, IL, the construction type of the building does not comply with NFPA 101 - 2000, 19.1.6.2. This deficiency could affect all in-patients in the facility, as well as those residents of the Long Term Care portion of the building and any staff and visitors present, because the lack of protection of the building structure from the effects of fire exposure can cause building collapse prior to evacuation.

Findings include:

A. The original 1934 building area is partially sprinklered and indicated to be of Type III (000) construction type and is two stories with a basement. The building construction type is not permitted under Table 19.1.6.2. Surveyor notes that the original building area is not otherwise separated from the Hospital by 2-hour rated construction on all floor levels to allow it to be classified as a separate non-healthcare occupancy building. The 2-hour separation is indicated to only exist on the 2nd floor.

UPDATE 4/9/15: At 8:30am on 4/9/15 it was observed that the 2-hour fire barrier wall and spray fireproofing for the steel floor structure constructed to separate the 1934 non-conforming building construction type from the hospital was complete, but fire stopping had yet to be installed as part of the corrective work and separation at the Lab area was yet to be completed. Surveyor notes that the 1/8/15 PoC indicates a May 15, 2015 completion date for this building separation.

B. The 1954 and 1974 building areas are indicated to be of Type II (000) construction type. These building areas are also two stories with a basement. The building construction type is not permitted under Table 19.1.6.2 without full sprinkler protection. The building areas are only partially sprinkler protected.

UPDATE 3/7/13: Pending the facility architect has not completed the "assessment". This was to have been completed by 1/31/13 and is now scheduled to be completed by 3/31/13.
Formal Documentation Preparation and submission to IDPH due 11/1/12 revised date 7/1/13
IDPH Review due 1/1/12 revised date 8/15/13
Bidding and award of project to take place on 3/1/13 revised to 10/1/13
Construction start date was anticipated 4/1/13 revised to 12/1/13
Completion date 6/30/13 revised to 12/1/14.

UPDATE 4/9/15: Sprinkler protection for the entire building is ongoing. Portions of the 2nd floor remain to be provided with sprinkler protection.

No Description Available

Tag No.: K0130

Based on random observation during the survey walk-through, document review, and staff interview, the facility is not in compliance with a series of Life Safety and other code requirements that are not documented under other K-Tags.

Findings include:

A. Due to the number, variety, and severity of the life safety deficiencies observed during the survey walk-through, the provider shall institute appropriate interim life safety measures until all cited deficiencies are corrected. The provider shall include, as an attachment to its Plan of Correction (PoC) and referenced therein, a detailed narrative and proposed schedule for all such measures. The narrative shall describe all measures to be implemented, as well as the frequency with which they are to be conducted, and shall indicate the manner in which the measures are to be documented. The narrative shall also include comments related to changes in the interim life safety measures to remain in place as work toward the completion of its PoC progresses.

No Description Available

Tag No.: K0147

Based on random observation during the survey walk-through, not all portions of the facility's electrical distribution system are in accordance with NFPA 70 1999.

Findings include:

A. Corrected 04/20/2012.
B. Corrected 04/20/2012.

C. During the facility walk-through on the afternoon of July 5th 2011 the emergency power breaker panels were observed with non-essential circuits powered from them with no separation of the 3 required branches of Critical, Life Safety and Equipment branches in accordance with 517-30 (b).

This deficiency could cause the overloading of the emergency generator causing failure which could result in patient injury.

UPDATE 3/7/13: Pending the facility architect has not completed the "assessment". This was to have been completed by 1/31/13 and is now scheduled to be completed by 3/31/13.
Formal Documentation Preparation and submission to IDPH due 11/1/12 revised date 7/1/13
IDPH Review due 1/1/12 revised date 8/15/13
Bidding and award of project to take place on 3/1/13 revised to 10/1/13
Construction start date was anticipated 4/1/13 revised to 12/1/13
Completion date 6/30/13 revised to 12/1/14.

UPDATE 4/9/15: The provider has implemented work in accordance with their 1/8/15 PoC to the 9/18/14 Recertification Survey. Work is currently scheduled to be completed by May 31, 2015.