505 SOUTH JOHN REDDITT DRIVE

LUFKIN, TX 75904

Sprinkler System - Installation

Tag No.: K0351

Based on observation the facility failed to provide adequate protection over an existing supply duct that is over 4' wide in the mechanical penthouse.

Sprinklers shall be installed under fixed obstructions over 4 ft. wide such as ducts, decks, open grate flooring, cutting tables, and overhead doors - NFPA 13, 2002: 8.5.5.3.1.

Operating Features - Other

Tag No.: K0700

Based on observation the facility failed to provide adequate letters indicating preferred customers status in case of emergency situation for water and fuel source.

Health Care Facilities Code - Other

Tag No.: K0900

Based on observation the facility failed to provide adequate signage for an existing low headroom location at mechanical penthouse.

HLR 133.132 (d)(2)(A)(xii)(III) Overhead clearance. Suspended tracks, rails, pipes, signs, lights, door closers, exit signs, and other fixtures that protrude into the path of normal traffic shall not be less than six feet eight inches above the finished floor.

Electrical Systems - Essential Electric Syste

Tag No.: K0917

Based on observation the facility failed to provide adequate identification to critical care outlets with non-removable labels-circuit and panel.