Tag No.: A0951
Based on a review of medical records, staff interviews, and review of hospital policies, it was determined that the hospital failed to ensure compliance with the policy entitled, "Surgical Counts", for relevant sample patient ID #2; and, the policies entitled "Tamponade Balloon Catheter" and "Report (Hand Off Communication)", for relevant sample patient ID # 3.
Findings are as follows:
1. A review of the hospital policy entitled, "Surgical Counts", section G, "Inaccurate Counts", states:
Under item #2, "Obtain order for X-ray".
Under item #3, "Document physician review of X-ray in the medical record".
A review of the medical record for patient ID #2 revealed the patient underwent a "Robot assisted hysterectomy, bilateral salpingo-oophorectomy, bilateral pelvic and periaortic lymph node dissection" on 8/25/10. An Occurrence Report submitted by the Circulating Nurse dated 8/25/10 revealed that during the procedure, the Scrub Technician noted that a "sponge had separated", resulting in a missing blue radiopaque string. The Surgeon was notified, and with visual inspection determined that this radiopaque string had been retained in the patient's abdomen, and "was able to find and remove it".
The patient was seen on 9/13/10 for an irritation of the wound site. When this persisted and there was a concern for cellulitis, a CT scan of the abdomen was performed on 9/27/10 and revealed a 4.8 x 1 cm (centimeter) fluid collection of the anterior wall superficial to the peritoneum, with a foreign body noted extending through the wall and into the peritoneal space. The patient returned to the Operating Room on 10/5/10 for a wound exploration. A Pathology Report dated 10/6/10 revealed "multiple fragments of light blue-red pieces ranging in measurement from 1.7 x 12.5 cm".
During an interview with the Scrub Technician on 10/28/10 at 12:10 PM, it was reported that when the Surgeon retrieved the radiopaque string with the initial surgery on 8/25/10, it had been measured against another similar sponge string, and "appeared to be the same in size". The surgical team "felt confident that all the radiopaque string had been removed", therefore an X-ray was not requested to confirm that the string had been retrieved in its entirety.
During an interview on 10/28/10 at 10:20 AM with the Surgeon, it was reported that the sponge had been delivered through the trocar during the surgery to "blot any bleeding" in order to provide visualization of the site. When the Surgeon was made aware by the Scrub Technician that a blue radiopaque string from a sponge utilized during the procedure was missing, a laparoscope was utilized and "this long blue string was visualized". It was removed, and the Surgeon then proceeded to do another "sweep" visualization to be sure all the string had been removed. The Surgeon stated, that at that point, "I had no doubt that I had gotten it all." With the string retrieved, it was determined that "an X-ray was not needed".
During an interview on 10/25/10 at 10:00 AM with the Risk Manager, it was reported that there has been no other occurrences regarding missing sponge strings. The hospital immediately changed the sponges used in the pelviscopy trays, and notified the manufacturer. They also put an action plan in place that included "if there is any question in regards to a product (sponge/equipment) used during a surgical procedure, an X-ray will be obtained prior to closure to confirm no retained foreign object", and "sponges will be unfolded and inspected prior to procedure".
Although the sponge count was correct, the hospital failed to ensure that all fragments from the radiopaque string had been accounted for by obtaining an X-ray.
2. A review of the hospital policy entitled "Tamponade Balloon Catheter" under "Purpose" states:
"In the event of a postpartum hemorrhage, the OB (Obstetrical) physician may insert a tamponade balloon catheter into the uterus in an effort to achieve hemostasis".
Under "Procedure: Assisting with Vaginal Placement of the Tamponade Balloon Catheter", it states, under item # 11:
"The vaginal canal may be packed with vaginal sponges if desired by provider.......Count the sponges prior to insertion and document in electronic record".
Under "Assisting with Removal of Tamponade Balloon Catheter", it states:
Under item #1: "Removal of the balloon catheter is performed by the physician within 24 hours of placement"; and,
Under bullet #2: "Remove and count vaginal sponges if placed (obtain X-ray if sponge count is not correct)."
A review of the hospital policy entitled, "Report (Hand Off Communication)", under "Purpose", states:
"To assure that adequate information is communicated to caregivers".
Under "Policy", item #1 states: "A caregiver to caregiver report (hand off communication) is given when a patient's care is transferred from one caregiver to another...."
Under Item #3, it states: "The report includes but is not limited to....Assessment..... equipment".
A review of the medical record for patient ID # 3 revealed a spontaneous vaginal delivery resulting in a viable female infant on 7/24/10. The patient was noted with postpartum bleeding despite administration of Pitocin, Misoprostol, and Hemabate. A decision was made to place a Bakri (tamponade) balloon to control bleeding.
Surgical documentation revealed that this was done under ultrasound guidance and "A Bakri balloon was guided until the tip reached the fundus of the uterus, and then it was held in place as a second provider inflated the balloon with fluid up to 300 ml (milliliters). After this, one roll of Kerlix was placed into the vagina to hold the Bakri Balloon in place." Further documentation by nursing revealed that the balloon was slowly deflated by two residents. At 10:00 PM on 7/25/10, a nurse's note revealed, "Bakri balloon out".
During an interview on 10/25/10 at 9:50 AM with the Risk Manager, it was reported that this patient presented to the clinic on 9/28/10 for complaints of a foul vaginal odor. During physician examination, the Kerlix roll was discovered and removed from the vaginal cavity. The patient was placed on prophylactic antibiotics.
During an interview with the Chief Resident on 10/26/10 at 12:35 PM, it was reported that after the Bakri Balloon had been placed, the Obstetrical team changed. During the team "handoff", which occurs twice a day, the teams meet to discuss the status of all patients being followed. In this case, it had not been communicated that there was a Kerlix sponge in place with the Bakri balloon. This resulted in the sponge not being removed and accounted for when the balloon fell out. The Chief Resident also reported that not all providers utilize sponges with the Bakri balloon procedure. The Kerlix sponge use is dependent on cervical dilation, and prevents the balloon from falling out as needed. It was reported that it is not unusual for the sponge to fall out when the balloon comes out.
During an interview on 10/27/10 at 10:30 AM with the Chief of Obstetrics, it was reported that this balloon has been utilized for approximately 2 years at the hospital, with no previous occurrences. The hospital utilizes this balloon approximately 25 times per year. The Medical Staff and Residents did attend a simulation in November of 2009, which included the management of postpartum hemorrhage and blood loss estimates at delivery, and the use of the Bakri balloon with a Power Point presentation.
During the interview with the Risk Manager, it was reported that the hospital has an action plan in place to standardize a documentation form for placement, care and removal of the Bakri balloon that includes clear documentation of use of sponges. In addition, standardization of communication handoff between residents is also planned.
It was determined that the hospital failed to ensure compliance with the Bakri balloon procedure relative to the counting of vaginal sponges when this balloon is removed.
The hospital also failed to follow their hospital policy and standard of practice related to adequate communication of information when care in handed off between caregivers.
Tag No.: A0955
Based on record review, staff interview, and review of the hospital policy entitled, "Informed Consent", it was determined that the hospital failed to properly execute the surgical informed consent, including the time that consent was obtained, for 5 of 8 relevant sample patients (ID #'s 10, 11, 12, 13 and 15).
Findings are as follows:
A review of the hospital policy entitled, "Informed Consent", states, under "Policy":
"The exact date and time at which consent was obtained must be indicated."
1. A review of the medical record for patient ID #10 revealed a surgical procedure on 10/18/10.
2. A review of the medical record for patient ID #11 revealed a surgical procedure on 10/8/10.
3. A review of the medical record for patient ID #12 revealed a surgical procedure on 10/19/10.
4. A review of the medical record for patient ID #13 revealed a surgical procedure on 10/8/10.
5. A review of the medical record for patient ID #15 revealed a surgical procedure on 8/2/10.
A review of the "Operative Consents" for all above patients revealed no documented times that these consents were obtained, in accordance with the hospital policy.
During an interview on 10/29/10 at approximately 1:30 PM with the both the Nurse Manager of Surgical Services and the Risk Manager, both were unable to provide evidence that the time that consents were obtained had been documented in the Informed Consents in accordance with hospital policy.