SURGICAL SERVICES

Tag No.: A0940

Based on observation, document review, and interview, facility failed to ensure proper temperature and humidity within accepted standards of practice.


Findings:


Observation of thermo-hygrometer showed out of control-band humidity reading in procedure room, 64%.


Interview with CST (25 July 2016 @ 10:50 AM) confirmed the hygrometer showed out of control-band humidity reading in procedure room, 64%.


Review of temperature and humidity logs, showed numerous out of control-band excursions of temperature and/or humidity in OR, Procedure rooms, and sterile supply.


Interviews with CEO (21 July 2016 @ 1:50 PM) and CNO (22 July 2016 @ 3:10 PM) confirmed temperature and humidity logs showed numerous out of control-band excursions for temperature and/or humidity in OR, Procedure rooms, and sterile supply.

MEDICAL STAFF - SELECTION CRITERIA

Tag No.: A0050

Based on document review and interview, facility failed to ensure that practitioners providing contracted services were required to be privileged.

Findings:

Review of bylaws indicated practitioners were required to be privileged but did not address privileging of contracted practitioners.

Review of contracted services with an anesthesia service and with a neurodiagnostics entity revealed that privileging was not addressed in either contract.

Interview with CEO (21 July 2016 @ 1:50 PM) confirmed that privileging of contracted practitioners was required and acknowledged that privileging was not addressed in the contracts nor in the bylaws for contracted provider services.

PATIENT RIGHTS: GRIEVANCES

Tag No.: A0118

Based on document review and interview, facility failed to have a comprehensive Complaint and Grievance policy.

Findings:

Review of policy 'Patient Complaint / Grievance - Flow Chart" did not:a) contain wording stating that a written or oral complaint, when requested, would be treated as a grievance.
b) contain wording about a written complaint must be treated as a grievance.
c) contain wording about complaints of abuse by staff results in staff being retasked until investigation has been completed.

Interview with CEO (20 July 2016 @ 1:30 PM) affirmed that 'Patient Complaint / Grievance - Flow Chart" did not contain wording about a written complaint necessitating being treated as a grievance; that any complaint, either written or oral, when requested, would be treated as a grievance; or that complaints of abuse by staff results in staff being retasked until investigation has been completed.

MAINTENANCE OF PHYSICAL PLANT

Tag No.: A0701

Based on observation and interview facility failed to maintain the physical plant in a manner to ensure safety and wellbeing of patients.

Finding:

Observation revealed two (2) sprinkler heads in procedure room #2 were missing escutcheons.

Interview with CST (25 July 2016 @ 10:50 AM) confirmed that two (2) escutcheons were missing from sprinkler heads in procedure room #2.

FACILITIES, SUPPLIES, EQUIPMENT MAINTENANCE

Tag No.: A0724

Based on observation and interview, facility failed to ensure that equipment was maintained at an acceptable level of safety and quality.

Findings:

Observation of Bair Hugger in procedure room 3 (store room) found it to be late for maintenance. Maintenance was due on 4/16. (Current date is 7/16)

Observation Dinamap #EI 31 in PACU; found it to be late for maintenance. Maintenance was due on 4/16. (Current date is 7/16)

Interview with CST (25 July 2016 @ 11:45 AM) confirmed that Dinamap #EI 31 in PACU and Bair Hugger in procedure room 3 were both due for maintenance. Maintenance for each was due on 4/16. (Current date is 7/16)

VENTILATION, LIGHT, TEMPERATURE CONTROLS

Tag No.: A0726

Based on observation, document review, and interview facility failed to ensure proper ventilation and temperature controls in appropriate areas.

Findings:

Observation of thermo-hygrometer showed out of control-band humidity reading in procedure room #1, 64%.

Interview with CST (25 July 2016 @ 10:50 AM) confirmed the hygrometer showed out of control-band humidity reading in procedure room #1, 64%.

Review of temperatures and humidity logs, showed numerous out of control-band excursions of temperature and/or humidity in OR, Procedure rooms, and sterile supply.

Interviews with CEO (21 July 2016 @ 1:50 PM) and CNO (22 July 2016 @ 3:10 PM) confirmed temperature and humidity logs showed numerous out of control-band excursions for temperature and/or humidity in OR, Procedure rooms, and sterile supply.

INFECTION CONTROL PROGRAM

Tag No.: A0749

Based on observation, document review, and interview, facility failed to mitigate risks contributing to healthcare-associated infections.

Findings:

Observation of thermo-hygrometer showed out of control-band humidity reading in procedure room, 64%.

Interview with CST (25 July 2016 @ 10:50 AM) confirmed the hygrometer showed out of control-band humidity reading in procedure room, 64%.

Review of temperature and humidity logs, showed numerous out of control-band excursions of temperature and/or humidity in OR, Procedure rooms, and sterile supply. Logs also showed numerous out of control-band temperature excursions for refrigerators and/or freezers in PACU, Med room, wound care, and patient kitchen.

Interview with CEO (21 July 2016 @ 1:50 PM) confirmed temperature logs showed numerous out of control-band excursions for temperature and/or humidity in OR, Procedure rooms, and sterile supply as well as for refrigerators and/or freezers in PACU, Med room, wound care, and patient kitchen.

Interview with CNO (22 July 2016 @ 3:10 PM) confirmed temperature and humidity logs showed numerous out of control-band excursions for temperature and/or humidity in OR, Procedure rooms, and sterile supply.

HISTORY AND PHYSICAL

Tag No.: A0952

Based on document review, and interview, facility failed to complete and document a medical history and physical no more than 30 days before or 24 hours after admission.


Finding:


Document review found patient #6 (one of twenty patient records) H & P was completed outside of the acceptable window.

H & P was dated 03/05/2016, date of service was 02/17/2016.


Interview with DON (25 July 2016 @ 3:01 PM) agreed that in one of twenty patient records H & P was completed outside of the acceptable window.

OPERATING ROOM REGISTER

Tag No.: A0958

Based on document review, and interview facility failed to maintain complete and up-to-date operating room register.


Findings


Document review of "SURGERY AND IMPLANT LOG" showed that:

a) 28 of 28 cases reviewed, log does not show inclusive or total time of the operation.

b) 28 of 28 cases reviewed, log does not show pre and post-op diagnosis.

c) 28 of 28 cases reviewed, log does not show anesthesia provider or other persons in attendance



Interview with DON (22 July 2016 @ 3:10 PM) confirmed that "SURGERY AND IMPLANT LOG" does not consistently show inclusive or total time of the operation, anesthesia provider, or the pre and post-op diagnosis.